Biotechnology Bulletin ›› 2018, Vol. 34 ›› Issue (8): 75-79.doi: 10.13560/j.cnki.biotech.bull.1985.1017-1084

Previous Articles     Next Articles

Antibodies Preparation and the Establishment of Chemiluminescence Immunoassay Detection Method for Cystain C Polypeptide

WANG Yun-long1,2,3, XU Rui-fang1, LI Yu-lin2,3, WANG Ji-chuang2,3, CHENG Chun-jie3, GAO Yu-hong3   

  1. 1. School of Life Sciences,Zhengzhou University,Zhengzhou 450001;
    2. Henan Bioengineering Technology Research Center,Zhengzhou 450001;
    3. Henan Engineering Laboratory for Biomarker Quantitative Detection,Zhengzhou 450001
  • Received:2017-12-18 Online:2018-08-26 Published:2018-09-04

Abstract: The objective of this work is to prepare cystain C antibody and establish the chemiluminescence detection method for the early diagnosis of renal damnification. Cystatin C monoclonal antibody was prepared through chemical synthesis of cystain C polypeptide,then competitive inhibition assay was used for preliminary screening of paired antibody,and chemiluminescence immunoassay detection method of cystain C was established. Totally,39 strains of cystatin C monoclonal antibody cells were obtained,and paired antibody was screened. By analyzing the linear range,minimum detection limit,precision,accuracy and other performance evaluation analysis,chemical luminescence detection method for detecting cystatin C was established,with linear range of 50--20 000 ng/mL and the minimum detection limit of 3 ng/mL. Conclusively,cystatin C paired antibody was prepared and chemiluminescence detection method was established successfully,which provides a new detection technique and method for early renal damnification.

Key words: cystatin C polypeptide, antibody paired, chemiluminescence