生物技术通报 ›› 2018, Vol. 34 ›› Issue (12): 77-83.doi: 10.13560/j.cnki.biotech.bull.1985.2018-0481

• 技术与方法 • 上一篇    下一篇

荧光定量PCR辐照食品鉴定技术的优化研究

刘绵学1, 2, 伏毅1, 2, 郑宇1, 王艳1, 2, 潘琳1, 李华1, 黄敏1, 2   

  1. 1. 四川省原子能研究院辐照保藏四川省重点实验室,成都 610101;
    2. 四川省原子能研究院生物技术研究所,成都 610101
  • 收稿日期:2018-05-24 出版日期:2018-12-26 发布日期:2018-12-24
  • 作者简介:刘绵学,男,博士,研究方向:分子生物学及辐照食品鉴定;E-mail:35837681@qq.com
  • 基金资助:
    四川省科技支撑计划项目(2016SZ0075,2016NZ0117),四川省应用基础项目(2017JY0242,2016JY0030,2018JY0531,18YYJC0284),四川省科技计划重点研发项目(2018GZ0003)

An Optimization on Identification of Irradiated Food by Fluorescence Quantitative PCR

LIU Mian-xue1, 2, FU Yi1, 2, ZHENG Yu1, WANG Yan1, 2, PAN Lin1, LI Hua1, HUANG Min1, 2   

  1. 1. Irradiation Preservation Key Laboratory of Sichuan Province,Sichuan Institute of Atomic Energy,Chengdu 610101;
    2. Department of Biotechnology,Sichuan Institute of Atomic Energy,Chengdu 610101
  • Received:2018-05-24 Published:2018-12-26 Online:2018-12-24

摘要: 为优化荧光定量PCR辐照食品鉴定技术方法,实现对假阳性样本的鉴别能力,降低模板量对鉴定结果的干扰效应。选择马铃薯作为研究材料,分别进行1-12 kGy的60Co辐照、加热和冻融处理,通过转座子引物筛选、检出限分析、DNA损伤特征比较和剂量曲线回归等方法,探究转座子引物的应用对马铃薯辐照鉴定结果的优化效果。结果显示,转座子引物组合LTR2-5\18S-8与LTR2-2\ACT-5可用于假阳性样本的鉴别,引物组合LTR2-5\18S-5可用于辐照鉴定。优化后方法中“初始循环数差值的指数函数-辐照剂量”标准曲线方程为Y=-0.1581X+ 4.611,线性关系R2为99.24%,优于“初始循环数差值-辐照剂量”的原有模式。通过转座子引物的应用与方法优化,增强了假阳性鉴别能力,消除了模板差异,增强了荧光定量PCR辐照鉴定技术的“剂量-效应关系”。

关键词: 辐照鉴定, 转座子, 马铃薯, DNA损伤

Abstract: Purpose of this study is to realize the identification ability of false positive samples and reduce template amount interference with real-time PCR results in optimizing irradiation food identification. Potato tuber was selected as material and treated with 60-Co in gradient dose(1-12 kGy),irradiating,heating and freezing-thawing respectively. Analysis methods,including primer screening of transposon,detection limit analysis,comparison of DNA damage characteristics and dose curve regression,were used to explore the optimized effects of transposon primers in potato irradiation identification. As results,transposon primer combination of LTR2-5\18S-8 and LTR2-2\ACT-5 can be used for the identification of false positive samples,and the combination of LTR2-5\18S-5 can be used for irradiation identification. In the optimized method,the irradiation dose was linear with the exponential function of the initial cycle difference with the standard equation Y=-0.1581X+ 4.611 and determination coefficient R2 99.24%,which was superior to the original initial cycle difference-irradiation dose model. In conclusion,by the optimized method,the dose-effect relationship with real-time PCR in irradiation identification was improved with less possibility of false positive,and the result interference caused by template amount was eliminated.

Key words: irradiation identification, transposon, potato, DNA damage